APP下载

卡尔费休库仑法微量水分测定仪示值误差的不确定度评定

2016-10-21严安陈会涛

价值工程 2016年9期

严安 陈会涛

摘要:为了获得卡尔费休库仑法微量水分测定仪示值误差的不确定度,通过将被校仪器测量值与标准物质标准值两者差值作为卡尔费休库仑法微量水分测定仪示值误差,依据JJG044-2008《卡尔·费休库仑法微量水分测定仪检定规程》、JJF1059.1-2012《测量不确定度评定与表示》等文献,评定出卡尔费休库仑法微量水分测定仪示值误差的不确定度。

Abstract: In order to obtain the uncertainty of indication error of Karl Fischer coulometric determination microscale water content determinator, the difference of the measured value of calibrated instruments and the standard values of standard material is taken as the indication error of Karl Fischer coulometric determination microscale water content determinator. According to JJG044-2008 "Vertification Regulation of Karl·Fischer Coulometric Determination Microscale Water Content Determinator", JJF1059.1-2012 "Evaluation and Expression of Measurement Uncertainty" and other literatures, the uncertainty of Karl Fischer coulometric determination microscale water content determinator is evaluated.

关键词:卡尔费休库仑法;水份测定仪;示值误差;不确定度

Key words: Karl Fischer coulometric determination;water content determinator;indication error;uncertainty

中图分类号:TH837 文献标识码:A 文章编号:1006-4311(2016)09-0227-03

0 引言

随着科学研究的发展和生产技术的进步,水分的定量分析已被列为各类物质理化分析的基本项目之一,作为各类物质的一项重要的质量指标。卡尔费休库仑法是其中一种既经济有准确的方法,被许多国家定为标准方法,用来校准其他方法和测量仪器。正确有效的对卡尔费休库仑法微量水分测定仪示值误差进行不确定度评定,对卡尔费休库仑法微量水分测定仪的使用及校准都具有重要的意义。

1 概述

①校准依据:JJG1044-2008《卡尔·费休库仑法微量水分测定仪检定规程》。②环境条件:室温:(10~30)℃,相对湿度:≤80%。③测量标准:液体水分标准物质GBW13512,其定值的标准值为1.000×10-2,扩展不确定度为0.013×10-2(k=2);液体水分标准物质GBW13513,其定值的标准值为0.1066×10-2,扩展不确定度为0.0031×10-2(k=2);液体水分标准物质GBW13514,其定值的标准值为0.0139×10-2,扩展不确定度为0.0012×10-2(k=2)。④被校对象:卡尔·费休水分测定仪。⑤校準过程:选取干燥的10μL进样针,用天平对其称重M1,抽取10μL液体水分标准物质GBW13513,再次称重M2,两次的差值M即为……