山东省医用CT及磁共振设备检定标准化管理现状与改进研究
2025-08-06刘洋刘再涛樊超刘增运侯铁栋宋飞霍太昌


Research on the Standardization Management Status and Improvement of Medical CT and Magnetic Resonance Equipment Calibration in Shandong Province
LIU Yang LIUZaitao FAN Chao LIU Zengyun HOU Tiedong SONG Fei HUO Taichang (Shandong Institute of Metrology)
Abstract:This study explores thecurent status ofcalibration management of medical CTand magnetic resonance equipmentin Shandong province.Byanalyzingtheroutineverificationdataof medicalmetrology equipmentin Shandong province in 2O24,it was found that the initial inspection passrates of medical CT machines and magnetic resonance equipment were 93.71% and 94.87% respectively.After reexamination,the pass rates increased to 99.84% and 99.63% (204号 respectively.TheimpactofCTvalueerrors,image noise,scanninglayerthicknessaccuracy,signal-to-noiseratioand uniformityonequipment quality isanalyzed,and improvement measures suchas strengthening equipment management, optimizingcalibrationprocesses,introducing new technologies,andestablishing provincialcalibration centers are proposed.By implementing improvement measures,it is beneficial to enhancing the equipment qualificationrateand management level, providing a solid guarantee for precision medicine.
Keywords: medical CT; magnetic resonance equipment; standardization of verification; quality assurance
0 引言
山东省作为我国人口第二大省及经济强省,其医疗设备市场庞大且增长迅速。随着国家惠民政策如“新农合”的实施,医用CT、磁共振等高端诊断设备的使用频率激增,公众对设备准确性的要求也日趋严格。为保障医疗质量,每年对全省此类医疗设备进行周期性检定,覆盖二级以上医疗机构。2024年,直接服务医疗卫生单位800余家,检定医用CT设备1903台,医用磁共振819台,包括GE、西门子、飞利浦、联影等主流品牌。当前检定工作仍面临标准化管理不足等挑战,本文旨在探讨山东省医用CT及磁共振设备检定标准化管理的现状,分析存在的问题,并提出改进策略,以进一步提升医疗设备检定水平,保障医疗质量与安全。
1山东省医用CT及磁共振设备检定标准化管理现状
在精准医疗的推进实践中,精确度量依赖于高精度的测量工具,医学计量仪器的性能直接影响医疗服务的质量。2024年,山东省对医学计量设备进行的检定中,发现部分设备存在运行故障。医用CT机检定的合格率为 93.71% ,经复检后合格率提升至99.84% ;磁共振设备初次检定合格率为 94.87% ,复检后合格率增至 99.63% ,大多数设备需经过调整和水模校准等操作才能达到合格标准。
图1和图2为2024年各地区的医用CT机及磁共振设备初次检查的合格情况。
对于初次检查未通过的设备,要求相关医疗机构暂停使用,并在完成修复和调试后重新申请检查,确保医疗服务的连续性不受影响,同时保证每台设备都能达到最佳工作状态。……