Ⅱ级生物安全柜下降气流流速测量结果不确定度评定
2021-08-20马骏


【摘要】 生物安全柜是微生物实验室的必需品,其不仅能对操作者提供保护措施,还能在进行细菌培养时提供无菌无尘的工作环境。为了保证操作人员在操作生物安全柜过程中不受细菌的侵害,生物安全柜的各项指标需要定期进行校准。本文对生物安全柜的下降气流流速测量结果进行不确定度评定,给出测量结果的不确定度。
【关键词】 生物安全柜;下降气流;重复性;不确定度
【DOI编码】 10.3969/j.issn.1674-4977.2021.04.012
Evaluation of Uncertainty in Measurement Results of Downflow Velocity of ClassⅡ Biosafety Cabinets
MA Jun
(Liaoning Equipment Manufacturing Vocational and Technical College,Shenyang 110161,China)
Abstract: The biosafety cabinet is a necessity of microbiological laboratory. It can not only provide protection measures for operators,but also provide a sterile and dust-free working environment for bacterial culture. In order to protect operators from bacteria,the indicators of biosafety cabinet need to be calibrated regularly. This article evaluated the uncertainty in measurement results of downflow velocity of biosafety cabinet,provided the uncertainty of measurement results.
Key words: biosafety cabinet;downflow;repeatability;uncertainty
测量Ⅱ级生物安全柜的下降气流流速,可以使用标准计量器具进行直接测量,且测量结果受温度、湿度、压力等环境因素的影响可以忽略不计。因此,本文主要分析测量过程中对测量结果影响较大的不确定度分量,并对其进行不确定评定。
1 测量方法
1.1 测量与评定依据
JJF 1815-2020《Ⅱ级生物安全柜校准规范》和JJF 1059.1-2012《测量不确定度评定与表示》。
1.2 环境条件
环境温度:(15~30) ℃;湿度:≤70% RH。
1.3 测量标准
风速仪,测量范围为(0~2) m/s,最大允许误差为±0.015 m/s。
1.4 被测对象
Ⅱ级生物安全柜的下降气流流速,下降气流平均流速应在(0.25~0.40) m/s之间。
1.5 测量过程
按照JJF 1815-2020《Ⅱ级生物安全柜校准规范》7.2要求,在生物安全柜的工作区上方高于前窗操作口上沿100 mm的水平面上确定测量点位置,然后多点测量穿过该平面的下降气流流速。测量區域边界与生物安全柜的内壁及前窗操作口的距离为150 mm,设置3排测量点,每排测量7个点,每个测量点测量3次,测量点等距均匀分布。用风速仪探头准确定位在各个测量点进行测量。
2 数学模型
[v=13ni=1nj=13vij] (1)
式中:[v]——平均下降气流流速,m/s;……p>
