腔内心电图定位技术在手臂输液港植入中的应用
2021-11-10李兰李雪红张艳钟远科詹雨薇
李兰 李雪红 张艳 钟远科 詹雨薇

【摘要】目的:觀察腔内心电图定位技术在手臂输液港中的应用。方法:选取2018年10月-2019年3月置入手臂输液港96例肿瘤患者,按随机数字表法分为观察组与对照组各48例。观察组在输液港植入过程中使用腔内心电图进行即时导管尖端定位,再进行X线定位;对照组常规输液港植入后行X线拍片定位,比较两种定位方法的准确率。结果:观察组患者44例(91.66%) 定位准确,对照组全部定位准确,2组定位准确率比较差异无统计学意义( x2=2.08,P>0.05) 。观察组中有44例(91.66%) P波振幅为QRS波50%~80%的心电图时即为导管尖端的最佳位置,胸部X线显示导管尖端位置在第5~7胸椎;有4例(8.3%) 患者P波形态异常,均未获得特异性P波或出现干扰波,胸部X线显示导管异位。结论:腔内心电图技术定位确定手臂输液港尖端位置准确率高,方法简单易行,可以在不增加患者痛苦和经济负担的情况下对导管的放置位置进行即时监测,调整导管位置,使穿刺和导管定位一次性完成,保证患者用药及时性,同时也减轻了护士工作量,避免反复复位带来的并发症及X线反复照射,有较高的临床应用价值。
【关键词】腔内心电图; 手臂输液港; 心电图特征性P波
【中图分类号】R472 【文献标识码】A 【DOI】10.12332/j.issn.2095-6525.2021.13.154
【Abstract】Objective:Application of intracavitary electrocardiographic localization technique in arm infusion port。Methods: 96 cases of tumor patients who needed to be put into the infusion port of arm from October 2018 to March 2019 were randomly divided into 48 cases of observation group and control group according to the random number table method. In the process of infusion port implantation, the observation group used the ECG to locate the tip of the immediate catheter and then conducted X-ray positioning verification. Control group of conventional infusion port implanted catheters to measure the length of the X-ray location of the patient. Compare the accuracy of the two positioning methods. Results:The patients in the observation group were placed with electrocardiogram and then verified with X lines. 44 cases(91.66%) were accurately positioned, all control groups were accurately positioned, and the positioning accuracy of the 2 groups was not statistically significant(x2 = 2.08, P> 0.05). In 44 cases(91.66%) of the observation group, the P-wave amplitude was 50 % to 80 % of the QRS wave electrocardiogram, which was the best position for the tip of the catheter. The chest X-ray showed that the tip of the catheter was located at the 5th to 7th thoracic vertebrae; In 4 cases(8.3%), there were no specific P waves or interference waves, and chest X-rays showed abnormal catheter. Conclusion:The method is simple and easy, so that the placement of the catheter can be monitored immediately without increasing the pain and financial burden of the patient, so as to adjust the position of the catheter, so that the puncture and catheter positioning can be completed at once. It is of high clinical application value to ensure the timely administration of medicine, and also reduce the nurse's workload, avoid the complications brought about by repeated reduction and X-ray exposure.
【Key words】Intracavity electrogram; Arm infusion port; Characteristic P wave of ECG
近年来,一种新的静脉输液技术手臂输液港开始应用,该技术操作简单,囊袋切口小,隧道短等,且埋植于上臂皮下,避免了血气胸、误穿动脉、导管夹闭综合征(Pinch-off)等并发症[1]。2016年INS指南首次提出手臂输液港可作为胸壁输液港替代选择[2],其适用于输注各种药物患者,以及各种血制品的输入及血样的采集等[3-4],近年在国内广泛开展。……