在HVI1000C检测中的问题及检查方法
2016-08-15李辉章张雪飞
李辉章 张雪飞
摘要:通过在棉花质量仪器化公证检验工作中的实践,分析了在HVI1000C仪器的测试数据的校准检查、根据反馈数据纠错校准等的一些问题,介绍了通过测试数据和反馈数据对实验室设备在仪器设备维护、环境温湿度、样品平衡、校准、校准检查、仪器操作等方面进行调整的方法。以提高棉花纤维检验实验室的工作质量。
关键词:HVI1000C;测试数据;反馈数据;问题与分析
中图分类号: S562. 文献标识码: A 文章编号:2095-3143(2016)04-0059-04
Abstract: Through the practice in the cotton quality inspection work, this paper analyzed some problems in the HVI1000C instrument test, such as the calibration check of the testing data, the error correction and calibration according to the feedback data, and so on. And it introduced the adjusting methods in the field of equipment maintenance, environmental temperature and humidity, sample balance, calibration, calibration check, instrument operation and so on through the testing data and feedback data. In order to improve the work quality of cotton fiber testing laboratory.
Key words: HVI1000C; Testing data; Feedback data; Problem and analysis
0引言
生产的棉花在纺织上应用,为了做到优棉优用就离不开纤维检验[1-3]。棉花的新品种选育、品种应用、栽培措施和生态布局要体现优质目标也离不开纤维检验[4-8],因此,棉花纤维检验的准确性十分重要。如今发表棉花纤维检验结果的论文较多[9-10],但对提高纤维检验结果准确性的论文鲜见,因此,作者介绍根据HVI1000C仪器的测试数据和反馈数据进行校准检查的方法,以提高棉花纤维检验数据的准确性。
在棉花质量仪器化公证检验工作中、服务器会收到大量的HVI1000C仪器的测试数据,校准检查数据;也会收到中国纤维检验局的抽检反馈数据。这些测试数据和反馈数据是对实验室设备进行调整的依据,也是评价实验室工作质量的依据。对这些测试数据和反馈数据进行全面、正确的分析,可以帮助我们发现实验室在仪器设备维护、环境温湿度、样品平衡、校准、校准检查、仪器操作等方面的存在问题,以便有针对性的纠正和预防,提高纤维检验的工作质量。……