医疗器械清洗应用ATP生物荧光检测法的检测效果分析
2016-07-16刘晓
刘 晓
医疗器械清洗应用ATP生物荧光检测法的检测效果分析
刘 晓
【摘要】目的 探究医疗器械清洗应用ATP生物荧光检测法的效果。方法 选取2015年3~9月由消毒供应中心提供的人工清洗和机械清洗后的医疗器械,基于ATP生物荧光检测法对其进行检测,对相对光单位值(RLU)进行测定,对器械清洗质量进行评价。结果 医疗器械总共400件,354件合格,其合格率为88.5%。基于不同清洁方式和不同类型下,对比清洗合格率,差异具有统计学意义(P<0.05)。结论 应用ATP生物荧光检测法对医疗器械清洗质量进行检测,效果优。
【关键词】医疗器械;ATP生物荧光检测法;检测效果
Objective Explore the effect of medical equipment cleaning application of ATP bioluminescence assay. Methods From March to September 2015,provided by the central sterile supply manual cleaning and mechanical cleaning of medical instruments,based on ATP bioluminescence assay for the detection and the relative light unit values(RLU) were measured,on equipment cleaning quality were evaluated. Results Medical equipment a total of 400 pieces,354 pieces of qualified the qualified rate was 88.5%. Based on the different cleaning methods and different types,the qualified rate of cleaning contrast,the difference with statistical significance (P<0.05). Conclusion The detection of quality of medical equipment cleaning and application of ATP bioluminescence detection method has significant effect.
【Key words】 Medical devices,ATP bioluminescence assay,Detection effect
清洗医疗器械是有效灭菌的关键所在,高质量清洗是提高灭菌有效率的保障[1]。目测法与细菌计数法是评价医疗器械清洗效果的主要手段,但其存在诸多弊端。基于此,ATP生物荧光检测法得到各大医院的关注与重视,为有效评定医疗器械清洗消毒质量提供依据[2]。为此,本院于2015年3~9月,采用ATP生物荧光检测法,取得一定成效,现将相关报道如下:
1 一般对象与方法
1.1 一般对象
2015年3~9月,每日随机抽取由消毒供应中心提供的人工清洗和机械清洗的医疗器械,总共包括400件,其中涉及普通手术器械、腔镜器械以及呼吸机管路等相关医疗器械。
1.2方法
1.2.1 清洗方法 (1)人工清洗:以《清洗消毒及灭菌技术操作规范》为指导,采用正确的方式清洗医疗器械。第一步,利用酶清洁剂将医疗器械进行浸泡和刷洗。第二步,通过流动的水、纯化水或者软水进行器械冲洗。……