一种医用自动计时血沉架的设计

2011-09-29陈桂涛

医疗卫生装备 2011年2期

陈桂涛

一种医用自动计时血沉架的设计

陈桂涛

目的:设计一种小型的适用于乡镇基层卫生院的医用自动计时血沉架。方法:整机主要由电源电路、计时电路、状态显示电路、语音提示电路4部分组成。结果:装上血沉管后,轻触开关自动接通该试管位的计时电路并开始计时;状态显示电路显示设备的正常工作状态,达到规定的时间后,状态显示电路和声音提示电路提示操作人员读取结果。结论:实验表明该设备计时准确、成本低、易于制作,所需配件均可在市场购得。按照该电路组装,无需调试可一次性成功,很适合目前乡镇卫生院、社区、个人门诊等基层医院使用。

自动计时血沉架;小型医用设备;工作原理;基层医院

AbstractObjectiveTo design an automatic self-timer medical blood sedimentation device for the primary hospital in villages and towns.MethodsThere were four circuit units in device,including the power supply circuit,self-timer circuit,display circuit and voice alarm circuit.ResultsTube's timing circuit unit was connected by touch switch after blood sedimentation tube was placed.The device's status was showed by the display circuit.The display circuit and voice alarm circuit could give an alarm when time was over.ConclusionThe device is reliable,accurate in counting time,easy to use and lower cost.The device's parts can be bought in any market,and can be assembled according to the diagram without any adjustment.It is suitable for such primary hospitals as the rural hospital,community hospital and clinic.[Chinese Medical Equipment Journal,2011,32(2):32-33]

Key wordsautomatic self-timer medical blood sedimentation device;small medical equipment;operation principle;primary hospital

1 引言

临床医学中的血沉分析实际上是指红细胞沉降率(ESR)(简称为血沉)的分析,ESR是常用的临床检验指标,在血液流变学测定中常作为红细胞聚集、红细胞表面电荷、红细胞电泳的通用指标[1]。目前,在一些基层医院,临床化验人员在检查血沉时仍普遍采用传统的魏氏手工测定方法[2]。工作中通常要对每只血沉管分别进行计时,这样不仅会使日常工作变得更加繁琐复杂,而且还会由于计时时间不准确直接影响到血沉读数的精度,形成误差,最终造成报告结果不准确。根据目前普遍存在的这种情况,我们设计制作了一种自动计时血沉架。其电路结构采用模块化设计,通过灵活的组合可同时对4、8、12、16、24只标准血沉试管分别计时(实际工作中我们采用对8只标准血沉管分别进行计时),当计时时间结束后,相应的试管位就会发光并发出“时间已到”的声音提示,及时提醒化验室操作人员观察血沉管的读数,基本上能够满足当前实际工作的需要。……