高精度砝码质量的测量及不确定度评定
2025-08-07康宝龙钟瑞麟
Abstract:High-precisionweightsareatype weighing instrument withhigh precisionand smallerrors,which are usuallyused inoccasions where therequirements forqualityorweightaccuracyareextremelyhigh,suchasscientific laboratories,pharmaceuticalindustrymetrologyandtesting,andotherfelds.Ensuringtheaccuracyandreliability high-precision weightvalues is greatsignificance inensuringproductqualityandimproving productperformanceand stability.Inthispaper,ameasurementmodel isestablishedforthe measurementprocessE2 standardweights,andthe verificationand extendeduncertaintyare evaluatedaccording tothe verification procedure JJG99-2022,Weights.The verificationresultsshowthattheagreedmass valuethe measuredweight meets therequirementstheregulations,and the extended uncertainty calculation results show that the measurement process is accurate and reliable.
Keywords:weight,mass,measurement uncertainty
1概述
1.1测量依据
JJG99—2022《砝码》。
1.2测量标准
E1等级标准砝码,约定质量 200g Ω
1.3被检对象
E2等级砝码,标称值 200g⨀"(2
1.4测量环境
砝码的检定应当在稳定的环境下进行,不允许有容易察觉的振动和气流,远离振动和磁场以及电离辐射。实验室必须配备相应准确度的温度计,湿度计和压力计用以测量环境条件。对于E2等级的砝码,实验室温度应控制在1 8∘C~23∘C ,每4h最大温度变化不得超过1 °C ,此次检定现场实验室环境条件如下:温度: 20.16∘C ;湿度: 44.2% RH;压力:1011.83hPa ;空气密度: 1.1975kg/m3 ;二氧化碳浓度: 345.1ppm 。
1.5测量过程
依据JJG99—2022《砝码》,首先进行外观检查,其次对砝码表面进行清洁。清洁过程中可使用无水乙醇等溶剂,清洁过程不得改变其表面特征,更不得损伤砝码。对于带有调整腔的砝码,不可使溶液浸入其中[2-3]。清洁后将砝码静置以使其达到稳定,对于E2级砝码,在使用无水乙醇清洗之后的稳定时间为24至48小时。一般情况下,推荐的稳定时间为24小时。最后,满足测量条件后,利用质量比较仪连续进行10次测量并记录,如表1所示,
2 测量模型
本次测量采用单次比较法,ABA循环,主标准器为E1等级的标准砝码,衡量仪器为微克天平。被检砝码的约定质量值与标准砝码的差值为:

式中: Vt ——被测砝码体积; Vr 标准砝码体积; ρ a——测量时实验室的实际空气密度;(20 ρ ——空气密度的参考值, 1.2kg/m3 ·ΔI. 天平的指示差; ΔIs ——由于灵敏度砝码引起的天平指示的改变; mcw 一 标准砝码的修正值。
3 测量不确定度评定
测量过程中的不确定度主要来源于: ① 多次重复测量引入的标准不确定度分量
,采用A类方法评定。……
