中性电极的热性能评估与标准符合性研究
2025-08-03李佳宁丁益王鹏飞张超
Study on Thermal Performance Evaluation and Standards Compliance of NeutralElectrodes
LI Jianing1DING Yi² WANG Pengfei³ ZHANG Chao³* (1.Centre Testing International Group Co.,Ltd.;2.ProSteri Testing Shanghai Co.,Ltd.; 3. KingPo Technology Development Co., Ltd.)
Abstract:With the widespreadapplicationof high-frequency electrosurgical devices inclinicalsurgery,thethermal performance and standards compliance of neutral electrodes,as critical safety components,are directly related to patient safety.BasedonthestandardGB97o6.202-2021,Medicalelectricalequipment—Part2-2:Particularrequirementsforthe basicsafetyandessentialperformanceof highfrequencysurgicalequipmentandhighfrequencysurgicalaccesories,this papersystematically interprets the heating mechanisms and standards compliance of neutral electrodes,and analyzes the scientificrationaleandimplementationrequirementsofthetestingcriteriaspecifiedinthestandard.Ithighlights theroot causes of discrepancies between the standard safety threshold (temperature rise ≤6∘C )and the clinical theoretical limit (temperature rise ≤12∘C ,evaluatestherationalityof graded test current levels( 350Ω-700mA )validatedbythethermal factormodel Θ=I2×t) ,and explores the risk characteristics ofhigh-current modes (gt;30A2S) . It aims to provide anin-depth interpretationof thecorerequirementsof the standard clauses,offercriticalguidance forneutral electrodemanufacturers inachieving standards compliance,optimize product designpathways,andensure the safetyand compliancein clinical application,thereby providing technical assurance for the safe implementation of high-frequency electrosurgical procedures.
Keywords: neutral electrode; thermal performance; high-current mode
0 引言
高频电外科设备是一种通过高频电流(通常200kHz~5MHz )作用于生物组织实现切割与凝血功能的医疗器械。其主要工作原理基于高频电流流经组织时因阻抗产生的焦耳热效应:当电流密度足够高时,局部组织温度迅速升高,导致细胞内水分汽化(切割)或蛋白质变性(凝血)。
以临床应用最广泛的单极模式为例,高频电流由发生器产生后,通过手术电极作用在目标组织,流经人体形成闭合回路,最终经中性电极(NeutralElectrode,临床上常称为负极板或回路电极)返回发生器,单极模式下电流路径如图1所示。中性电极作为单极模式下电流回路的必要组件,其核心功能是通过有效接触面积分散高频电流,将电流密度严格限制在安全阈值内,然而,临床数据显示,当电极贴合不良、材料老化或重复使用导致的接触面积下降,局部电流密度可能超过临界值,最终造成患者皮肤灼伤。
为了验证中性电极的热性能,GB9706.202—2021(修改采用IEC60601-2-2:2017)通过条款201.15.101.5明确规定:按使用说明书在正常使用状态下应用时,中性电极不应使患者接触部位受到热损伤风险]。该试验通过人类皮肤,或电热等效代用品或试验装置来验证其温升性能,确保测量点温升不超过 6∘C 。相较于在活体试验,代用品/试验装置的应用不仅规避了伦理争议与受试者安全风险,还可提高测试结果的重现性与可比性。

1 标准要求及分析
临床研究表明,当人体皮肤接触到灼热的固体表面时,可能造成灼伤。……