热释光个人剂量监测系统及其质量控制*
2016-07-23秦永春余宁乐王进王福如马加一陈维史晓东
秦永春余宁乐王 进王福如马加一陈 维史晓东
热释光个人剂量监测系统及其质量控制*
秦永春①余宁乐①王 进①王福如①马加一①陈 维①史晓东①
[摘要]目的:采取系列热释光个人剂量测量(TLD)系统质量控制措施,保证热释光个人剂量监测系统稳定可靠。方法:制定质量控制体系文件,严格规范个人剂量检测流程,对个人剂量测量系统采取定期维护保养、检定、期间核查及长期稳定性监测等质量控制措施。结果:通过对热释光个人剂量测量系统采取质量控制,在2014年的全国放射卫生技术服务机构个人剂量比对中取得了较好的成绩,参比的剂量计单项性能|Pi|≤0.1,综合性能|B|+S≤0.1,比对结果为优秀。结论:对热释光个人剂量测量系统采取的质量控制措施有效可行,个人剂量监测系统基本稳定可靠。
[关键词]个人剂量监测;热释光剂量计;质量控制

秦永春,女,(1981- ),硕士,主管医师。江苏省疾病预防控制中心,从事放射防护检测工作。
China Medical Equipment,2016,13(5):23-25.
[Abstract] Objective: To guarantee the stable and reliable performance of TLD individual monitoring system through a series of quality control measures. Methods: To establish the documents of quality control system and standard measurement procedure for TLD. The measures of maintenance, calibration and intermediate check of the TLD monitoring system were taken periodically. Results: Compared with the national individual monitoring, TLD system for individual monitoring achieved good results in 2014 with the |Pi|≤0.1, |B|+S≤0.1.
Conclusions: Quality control methods are taken effectively and performance of individual monitoring system is stable and reliable.
[First-author’s address] Jiangsu Provincial Center for Disease Control and Prevention, Nanjing, Jiangsu, 210009, China.
放射工作人员个人剂量监测是放射防护工作的重要组成部分,也是放射工作人员职业健康管理的重要内容[1]。为使监测结果具有可靠性和可比性,必须对监测进行质量控制,为此,本研究重点报告实验室热释光个人剂量监测系统及采取的质量控制措施,以保证热释光个人剂量监测系统的稳定可靠[2]。
1 热释光个人剂量测量系统
1.1 测量系统概述
热释光个人剂量测量(thermoluminescence dosimetry,TLD)系统使用的仪器有Harshaw 5500 和Harshaw 3500型热释光读出器(美国热电公司生产),TLD.LAB-01型退火炉(德国工业高炉和热能技术有限公司),FJ 417型照射器(北京核仪器厂)。……