自动管电流调制技术在颈部CT检查中的图像质量及辐射剂量控制中的价值
2017-05-30卓炼石
卓炼石

摘要:目的:探究自动管电流调制技术(ATCM)在颈部CT检查中的图像质量及辐射剂量控制中的价值。方法:选择本院于2014年12月-2016年11月间收治的87例颈部CT检查患者为研究主体。划分为A组和B组,分别是44例与43例。结果:A组的上颈部图像质量评分低于B组,下颈部图像质量评分高于B组,对比差异明显,有统计学意义(P<0 05),且87例患者的图像质量均在诊断要求范围内。A组的上颈部客观噪声水平高于B组,下颈部客观噪声水平低于B组,对比差异显著,有统计学意义(P<0 05)。B组的管电流大小并无明显变化,A组的上、下颈部管电流对比,有差异,有统计学意义(P<0 05)。A组的CTDIvol(容积CT剂量指数)低于B组,对比差异显著,有统计学意义(P<0 05)。结论:ATCM技术与固定管电流技术相比,可以明显控制患者颈部CT检查中的辐射剂量,且图像质量较佳,值得推广。
关键词:自动管电流调制技术;颈部CT;图像质量;辐射剂量
中图分类号:R445 3 文献标识码:B DOI:10 3969/j.issn.1001-0270.2017.04.13
The Value of Automatic Tube Current Modulation Technique in Image Quality and
Radiation Dose Control of Neck CT
ZHUO Lian-shi
(Anhui Navy Anqing Hospital, Anqing 246000, China)
Abstract: Objective: To explore the value of automatic tube current modulation (ATCM) technology in the image quality and radiation dose control of neck CT examination Methods: 87 cases of cervical CT patients admitted to our hospital from December 2014 to November 2016 were selected as research subjects The patients were divided into A group and B group, 44 cases and 43 cases, respectively The image quality and radiation dose were compared Results: The image quality score of A group was lower than that B group, the image quality score of lower neck was higher than that B group, and the difference was statistically significant(P<0 05) The image quality of 87 patients was within the range of diagnostic requirements The A group had higher level of noise on the upper neck than in the B group, and the lower cervical noise level was lower than that in the B group, and the difference was statistically significant(P<0 05) There was no significant change in the tube current in the B group, and there was a significant difference between the upper and lower cervical canal currents in the A group (P<0 05) The CTDIvol (volume CT dose index) of A group was lower than that B group, and the difference was statistically significant(P<0 05) Conclusion: Compared with the fixed tube current technique, ATCM technique can obviously control the radiation dose in patients with cervical CT examination, and the image quality is better, so it is worth popularizing
Key Words: Automatic Tube Current Modulation; Neck CT; Image Quality; Radiation Dose
頸部CT检查具有扫描时间偏短、速度较快等优势,被广泛用于临床颈部疾病的治疗中[1]。CT检查的图像分辨率由射线的剂量值决定,所以常通过加大辐射量的方式,获取更为清晰的图像[2]。本文旨在分析ATCM技术在颈部CT检查中的图像质量及辐射剂量控制中的价值,详细如下:
1 资料与方法
1 1 一般资料
选择本院于2014年12月-2016年11月间收治的87例颈部CT检查患者为研究主体。划分为A组和B组,分别是44例与43例。A组中,男26例,女18例;年龄范围是31-75岁,平均(55 24±3 11)岁;体重范围是47-62kg,平均(58 21±3 61)kg。B组中,男25例,女18例;年龄范围是33-74岁,平均(55 12±3 08)岁;体重范围是48-67kg,平均(58 27±3 17)kg。对比以上数据,差异不明显,无统计学意义(P>0 05),可比较。
1 2 方法
两组均利用多层螺旋CT进行颈部检查。A组给予ATCM技术处理,且NI(噪声指数)为10。管电压为120kv,管电流的调制范围75至500mA,Pitch为0 984,探测器的准直为64×0 625mm,机架转速为0 6s/圈,DFOV为20至25cm,SFOV为32cm。
1 3 观察指标
图像质量评价:将图像传至处理工作站,并有效去除图像表面的技术信息。将上颈部与下颈部的平扫图像作为质量评估主体,所有操作由1名医生+1名技师完成。
图像质量诊断要求:通过5分法进行评估,不能接受记为1分,平均以下记为2分,能够接受记为3分,平均以上记为4分,可很好接受记为5分。……p>