全自动消解—火焰原子吸收法测定土壤中的重金属
2016-05-14贺忠翔
贺忠翔



摘要:通过实验优化,建立了土壤样品全自动前处理消解程序,并用连续光源火焰原子吸收法同时测定土壤标准样品中的铜、铅和镍,取得了满意的结果。该消解方法教传统的方法有省时,加酸都能自动完成,减少人为操作带来的误差,精密度好,准确度更高。
关键词:全自动样品处理系统;连续光源火焰原子吸收法;铜、铅和镍
中图分类号:X833 文献标识码:A 文章编号:2095-672X(2016)05-0049-04
DOI:10.16647/j.cnki.cn15-1369/X.2016.05.011
Determination of heavy metals in soil by Automatic dugestion-Flame
atomic absorption
He Zhongxiang
( Inner Mongolia Environmental Monitoring Centre,Hohhot 010011)
Abstract:A method was established for the determination of some metal elements(Copper、Lead and Nickel) in soil standard by DEENA Automated Sample Preparation System-High Resolution Continuum Source Atomic Absorbsion Spectrometry((HS—CS—AAS).The recommended method can complete several elements' analysis quickly and effectively and can be used widely.Through analysis,accurate and reliable results can be got.
Key words:DEENA Automated sample preparation dystem;High Resolution Continuum Source Atomic Absorbsion Spectrometer;Copper、Lead and Nickel
引言
对于土壤样品的测试,前处理是非常关键的,前处理的好坏直接影响最终结果的准确性。常规的土壤前处理方法有:电热板法,微波消解法等。电热板法是最传统的方法,但是对于处理大批量的样品,需耗费大量的人力物力,从称样到加酸到最后定容全程都需有人看管;微波消解法是一种高效的样品前处理方式,因其是一种密闭样品前处理体系,固有用酸量少,节约资源,减少污染,正越来越被广泛地应用于各种样品的前处理,特别是用于处理土壤样品,可大大提高前处理的效率[1][2]。但微波消解同样需转移到就坩埚中,于电热板上赶酸,只能比电热板稍节省时间;而我们建立的土壤样品DEENA全自动前处理方法,只需把样品称量好,然后置于样品架上,运行程序仪器就会自动完成加酸、加热、赶酸、冷却及定容的全过程,而且加热面积大,大大地节约了时间,节省了人力。三种消解方法对比数据如下表,从表中可知DEENA全自动前处理方法的优点是显而易见的,特别是用于大批量样品的处理。……