MF1型流量显示仪的一种系统校准方法
2019-06-18崔艳文
崔艳文


摘 要:通过对MF1型流量显示仪工作原理及使用中测量误差产生原因的分析,提出一种便于操作的系统校准方法。调整显示仪闸门时间,使其与流量计仪表系数相匹配来修正测量误差。同时校正显示仪模拟输出信号,实现流量计、显示仪和计算机(或其他测量设备)的整体系统校准。最后通过试验验证该校准方法的可行性及显示仪系统校准的必要性。
关键词:流量显示仪 系统校准 仪表系数 闸门时间 模拟输出
中图分类号:TH814 文献标识码:A 文章编号:1672-3791(2019)03(b)-0066-04
Absrtact: Based on the analysis of the working principle of MF1 type flow display instrument and the cause of measurement error in use, a convenient system calibration method is put forward. Adjust the gate-time of the display device so that it matches the meter coefficient of the Flowmeter to correct the measurement error. At the same time, the analog output signal of the display instrument is corrected, and the whole system calibration of the Flowmeter, the display instrument and the computer (or other measuring equipment) is realized. Finally, the feasibility of calibration method and the necessity of system calibration of display instrument are verified by experiments.
Key Words: Flow display instrument; System calibration; Instrument coefficient; Gate-time; Analog output
MF1型流量顯示仪(以下简称显示仪)是德国威仕公司(VSE)生产的一种脉冲输入型流量显示仪表,与该公司VS型流量传感器(以下简称流量计)配套使用,作为流量测试系统的主体。一般在实际工作中,显示仪和流量计是分开单独校准,使用时按量程匹配的原则直接连接进行测量。由于没考虑到流量计性能变化的影响,这样的测量结果显然是不准确的,存在测量误差。配套使用的流量测试系统应进行配套校准才能保证测量的准确度。但是,配套校准的方法受条件限制,不易实现。那么怎样才能解决由于显示仪和流量计分开校准而在实际应用中造成的测量误差问题,达到配套校准的效果。该文提出一种便于操作、效率高,在实验室即可完成的系统校准方法。
1 显示仪的工作原理
MF1型流量显示仪原理框图如图1所示。显示仪采集与之配套流量计输出的两路相位差90°脉冲信号,将其转换为4倍脉冲频率。……