Exposure dose calculation of digital tomosynthesis and in comparison with other chest examinations
2017-02-21CN
, , , , *, CN
(1.Department of Radiology, Nanfang Hospital of Southern Medical University, Guangzhou 510515, China;2.Department of Radiology, Nanhai Hospital, Southern Medical University, Foshan 528200, China)
·方法技术学·
Exposure dose calculation of digital tomosynthesis and in comparison with other chest examinations
MALe1,ZHUMingxin2,QINGenggeng1,HEZilong1,CAIYuxing1*,CHENWeiguo1
(1.DepartmentofRadiology,NanfangHospitalofSouthernMedicalUniversity,Guangzhou510515,China;2.DepartmentofRadiology,NanhaiHospital,SouthernMedicalUniversity,Foshan528200,China)
Objective To propose a simple method of calculating exposure dose in chest tomosynthesis (DTS) whose validity was verified, and to compare the expose paremeters of DTS with other chest examinations of digital radiology (DR), dual energy subtraction (DES), low-dose CT (LDCT) and conventional CT. Methods Thirty subjects were recruited for the DTS examination. Based on the exposure parameters from scout view, a calculation method and two correction factors were proposed and validated for air kerma and dose-area product (DAP) calculation. Besides, 30 subjects participated in the DR examination, 30 for DES examination, 30 for LDCT examination and another 30 for CT examination. The difference of effective dose (ED) among five chest examinations was investigated. Results Two correction factors of 0.97 and 0.93 for air kerma and DAP were determined, respectively. Small average differences of (0.5±0.4)% and (0.3±0.3)% were found between estimated values and actual values. ED of DR, DES, DTS, LDCT, CT had statistical significance (F=1 148.09,P<0.01), and besides the ED of DR and DES had no difference, there were significant differences between each other two imaging methods (allP<0.05). Conclusion A simple method to estimate DTS exposure dose is proposed and validated with no need to superimpose the dose from each image. DTS has lower exposure dose when compared with LDCT and CT, and provides three-dimensional imaging for obtaining more information when compared with DR and DES.
Thorax; Tomosynthesis; Radiation dosage
数字X线摄影(digital radiography, DR)是胸部疾病检查最常用的影像学手段,具有成像速度快、辐射剂量低、价格便宜等优点。但由于骨组织对肺组织的重叠遮挡,肺内小结节易被漏诊。胸部双能量减影(dual energy subtraction, DES)通过高千伏和低千伏连续两次曝光,可将骨组织和肺组织分离,有效去除重叠的肋骨等高密度骨质对肺小结节的影响[1]。数字断层融合(digital tomosynthesis, DTS)作为一种低剂量的胸部断层成像方式,通过有限角度的连续、快速曝光,可重建出一系列断层图像[2-3]。相比DR和DES,胸部DTS可避免深在部位和复杂部位的组织重叠,减少呼吸运动所造成的图像伪影[4]。研究[4-5]证实,胸部DTS可提高病灶的检出率和诊断率,为临床治疗提供重要影像资料。目前,CT仍是胸部疾病检查的金标准,但由于其费用高、辐射剂量大等缺点,极大地限制了其在临床的广泛应用[6]。低剂量CT(low-dose computed tomography, LDCT)在保证图像质量的前提下,通过降低X线球管电流强度可达到降低辐射剂量的目的[7],但临床尚未广泛应用。Kumar等[8]以CT作为金标准,发现DES诊断肺内小结节的敏感度和特异度均高于DR,而DTS则显著优于DR(P<0.003),但该研究未对辐射剂量进行分析。
空气比释动能、剂量面积乘积(dose-area product, DAP)和有效剂量(effective dose, ED)常用来评价患者在X线检查过程中所接受的辐射剂量[9-10]。但在计算胸部DTS的空气比释动能和DAP时,常需对不同角度拍摄的图像剂量(系统默认60张)进行叠加,得到最后的辐射总量[9]。本文在基于DTS首张定位片曝光参数的基础上,提出一种计算空气比释动能和DAP的简单方法;同时提出两个校正因子对剂量参数进行调整,并对该计算方法和两个校正因子进行验证,将胸部DTS与DR、DES、LDCT、CT成像方式的ED进行对比。……
猜你喜欢
杂志排行
中国介入影像与治疗学的其它文章
- Ultrasonography in diagnosis of right hand deficiency and right upper limb most deficiency in fetuses: Two cases report 超声诊断胎儿右手缺如及右上肢大部分缺如2例
- Combined interventional therapies for hepatocellular carcinoma associated with gastroesophageal varices: Case report 多种介入疗法联合治疗胃底食管静脉曲张并发原发性肝癌1例
- Inhibitory effect of low intensity ultrasound combined with HSV1-TK/GCV gene carried by cationic microbubbles on high-intensity focused ultrasound incompleted ablation of hepatocellular carcinoma cells
- Kawasaki disease with left coronary artery mild expansion and thrombosis:Case report 川崎病合并左冠状动脉轻度扩张及血栓形成1例
- Feasibility of dorsal scapular nerve block with ultrasound guidance
- Prenatal echocardiographic diagnosis of fetal right-sided aortic arch and right-sided ductus
