IMAR对起搏器植入患者胸部CT扫描的应用价值

2016-10-18曾雨薇孙静坤张黎王国树

中国医学计算机成像杂志 2016年4期

曾雨薇孙静坤张 黎王国树

IMAR对起搏器植入患者胸部CT扫描的应用价值

曾雨薇1孙静坤2张 黎3王国树2

目的:通过仿真胸部体模对伪影区域内肺结节及肺组织显示效果的分析,评价IMAR算法对心脏起搏器植入患者CT扫描的应用价值。方法:在胸部仿真体模内置入模拟肺结节60个,分别在心脏起搏器放置前后进行胸部CT扫描(常规剂量及低剂量扫描),图像分别采用无IMAR和有IMAR两种方式进行重建。比较肺结节的检出率,测量模拟肺结节CT值与其标称CT值的差值,并比较图像的噪声值、SNR、CNR值,并对主观图像质量评分。结果:低剂量胸部CT扫描时, 使用IMRA检出结节数目较多;使用IMAR算法磨玻璃肺结节的测量CT值与标称CT值的差值较小。使用IMAR重建时图像SNR及CNR值均较高。IMAR重建图像质量主观评分高于未使用IMAR。结论:IMAR算法能减轻心脏起搏器金属伪影,提高低剂量胸部CT对伪影区域结节的检出率,提高伪影区域磨玻璃肺结节的显示效果,提高伪影区域图像质量。

体层摄影术, X线计算机;金属伪影;重建算法;图像质量

【Abstract】Purpose: To evaluate the application values of IMAR(iterative metal artifact reduction)algorithm on CT scanning of cardiac pacemaker implanted patients. Methods: Sixty simulated pulmonary nodules were inserted into the anthropomorphic phantom. The phantom was scanned with and without the pacemaker respectively (low-dose CT and regular dose CT scan respectively). Images were reconstructed with and without IMAR respectively. The sensitivities in detection of pulmonary nodules and their deviation of CT values were compared between the two reconstructed methods. On the regular dose CT scan images, the image noise, SNR and CNR were calculated. And the subjective image quality scores were evaluated by two radiologists. Results:All the pulmonary nodules were detected by using the low-dose CT scan with IMRA reconstruction algorithm; the simulated pulmonary nodules with diameter of 3mm and CT value of 800HU were not detected without the use of IMRA algorithm. For the pulmonary nodules with a standard CT number of -800HU and -650HU, the deviation ofCT values were significantly smaller with IMAR algorithm than that without it. The SNR and CNR were higher when using IMAR. The subjective image quality scores of the images reconstructed using IMAR were higher than that without IMAR algorithm. Conclusion: IMAR algorithm can significantly reduce mental artifact of cardiac pacemaker, and improve the sensitivities in detection of pulmonary nodules with low-dose CT scan, the display effect of the nodules with ground-glass opacity (GGO), and image quality of the areas with artifacts.

【Key words】Tomography, X-ray computed; Metal artifact; Reconstruction algorithm; Image quality

很多心律失常患者需要使用心脏起搏器来维持正常心率,迄今为止,心脏起搏器是治疗心动过缓的唯一手段,然而安装了心脏起搏器的患者在进行常规CT检查时图像中会有伪影产生。由于患者体内含有金属等高密度物质,投影数据重建后图像中会出现金属伪影,这些伪影严重影响了图像的质量,给医生的诊断带来了很大困难[1]。目前尚无临床实用的消除金属植入物伪影的成像参数标准和影像诊断标准[2]。……

登录APP查看全文